Thermocouples for temperature measurement up to 1,700 °C

Thermocouples generate a voltage directly dependent on temperature. They are particularly suited for high temperatures up to 1,700 °C. Instrument designs from mineral-insulated sheathed cable are very resistant against extremely high vibration loads (depending on instrument model, sensor element and wetted materials). For thermocouples, the accuracy classes 1 and 2 apply (standard and special). They are available with a tolerance value in accordance with IEC 60581 / ASTM E230.

Sales contact

WIKA Instruments Namibia (Pty.) Ltd.

Tel.:
+26 4 61238811
Fax:
+26 4 61233403
sales@wika.com.na

FAQ

What are the response times of the various measuring inserts?

The measurement of the response time is carried out in flowing water in accordance with DIN EN 60751 Section 4.3.3 and VDI / VDE 3522.

v=0,4m/s T start = approx. 20°C  T end = approx. 30°C
 
6mm 1xTyp K, isolated: T50= 4,0sec. ... More

What is the permissible minimum bending radius for an MI cable?

VDI/VDE 3511 Sheet 2 recommends a radius of curvature R of ≥ 5 x D (D=external diameter of the MI cable), some manufacturers of MI cable even give ≥ 3 x D as the minimum bending radius.